The model presented in this work is realistic in terms of anatomical dimensions and mean physiological values in a healthy-adult patient, going beyond the patient-specific model limits. The minute volume used for both the stationary and transient CFD simulations is 6 L/min, representing a rest breathing condition with a sinusoidal pattern (T = 4 s). The regional deposition efficiency of particle about 1, 3, 7, 10 and 25 µm has been compared to in vivo data and theoretical models up to the fourth generation in the stationary case. In transient case, deposition efficiency of 7 µm particle during inspiration phase shows a specific relationship between the site of deposition and injection time, at pulmonary level regions.

Il presente lavoro fornisce un modello 3D di sistema respiratorio non paziente specifico, ma realistico, rispondente in tutte le sue dimensioni e prestazioni ai valori medi fisiologici che caratterizzano l’apparato di un paziente sano adulto. Il contesto del progetto ha determinato la scelta di un volume minuto respiratorio pari a 6 L/min, corrispondenti ad uno stato di respiro tranquillo. Le simulazioni CFD sono state svolte in condizioni sia stazionarie che tempo-dipendenti, rappresentando un ciclo respiratorio sinusoidale di 4 s. La deposizione di aerosol di diametro pari a 1, 3, 7, 10 e 25 µm è stata analizzata e validata fino alla 4^ generazione per il caso stazionario tramite la comparazione dei risultati in termini di efficienza di deposizione regionale con risultati in vivo e modelli teorici presenti in letteratura. In stato transitorio è stata analizzata la deposizione di aerosol di diametro pari a 7 µm, che mostra una relazione fra luogo di deposizione e tempo di ingresso specifica per ogni lobo polmonare.

Progettazione e validazione di un modello 3D realistico di sistema respiratorio per l'analisi della deposizione di aerosol, per applicazioni terapeutiche termali e studi ambientali

LESMA, GIANLUCA
2015/2016

Abstract

The model presented in this work is realistic in terms of anatomical dimensions and mean physiological values in a healthy-adult patient, going beyond the patient-specific model limits. The minute volume used for both the stationary and transient CFD simulations is 6 L/min, representing a rest breathing condition with a sinusoidal pattern (T = 4 s). The regional deposition efficiency of particle about 1, 3, 7, 10 and 25 µm has been compared to in vivo data and theoretical models up to the fourth generation in the stationary case. In transient case, deposition efficiency of 7 µm particle during inspiration phase shows a specific relationship between the site of deposition and injection time, at pulmonary level regions.
BIANCHI, ELENA
CORTELEZZI, LUCA
ING - Scuola di Ingegneria Industriale e dell'Informazione
28-apr-2017
2015/2016
Il presente lavoro fornisce un modello 3D di sistema respiratorio non paziente specifico, ma realistico, rispondente in tutte le sue dimensioni e prestazioni ai valori medi fisiologici che caratterizzano l’apparato di un paziente sano adulto. Il contesto del progetto ha determinato la scelta di un volume minuto respiratorio pari a 6 L/min, corrispondenti ad uno stato di respiro tranquillo. Le simulazioni CFD sono state svolte in condizioni sia stazionarie che tempo-dipendenti, rappresentando un ciclo respiratorio sinusoidale di 4 s. La deposizione di aerosol di diametro pari a 1, 3, 7, 10 e 25 µm è stata analizzata e validata fino alla 4^ generazione per il caso stazionario tramite la comparazione dei risultati in termini di efficienza di deposizione regionale con risultati in vivo e modelli teorici presenti in letteratura. In stato transitorio è stata analizzata la deposizione di aerosol di diametro pari a 7 µm, che mostra una relazione fra luogo di deposizione e tempo di ingresso specifica per ogni lobo polmonare.
Tesi di laurea Magistrale
File allegati
File Dimensione Formato  
2017_4_Lesma.pdf

accessibile in internet solo dagli utenti autorizzati

Descrizione: PROGETTAZIONE E VALIDAZIONE DI UN MODELLO 3D REALISTICO DI SISTEMA RESPIRATORIO PER L'ANALISI DELLA DEPOSIZIONE DI AEROSOL, PER APPLICAZIONI TERAPEUTICHE TERMALI E STUDI AMBIENTALI
Dimensione 12.07 MB
Formato Adobe PDF
12.07 MB Adobe PDF   Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/133429