This work aims to study the behaviour of the new type of socket designed, built and marketed by the company ITOP S.p.A. Officine Ortopediche: the hybrid sub-ischial socket. Using the Abaqus/CAE 6.14 software, the finite element analysis is carried out to calculate the stresses and deformations on the stump's external surface following the procedure of the donning of the prosthesis and the bipodalic stance in orthostatic posture. To verify the prosthesis's actual comfort, the stresses obtained in terms of contact pressure and shear stresses must fall within a range of values considered not harmful to the patient. This thesis focuses on studying the modelling of the materials constituting the stump and on the contribution of the liner compression on the stump.

Il presente lavoro si pone come obiettivo lo studio del comportamento della nuova tipologia di invasatura progettata, realizzata e commercializzata dall’azienda ITOP S.p.A. Officine Ortopediche: l’invasatura sub-ischiatica ibrida. Tramite l’utilizzo del software Abaqus/CAE 6.14 si conduce l’analisi agli elementi finiti, finalizzata al calcolo degli sforzi e delle deformazioni sulla superficie esterna del moncone sia al termine della calzata della protesi sia in posizione ortostatica con appoggio bipodalico. Al fine di verificare l’effettivo comfort della protesi, gli sforzi ottenuti in termini di pressione di contatto e di sforzi di taglio devono ricadere all’interno di un range di valori considerato non dannoso per il paziente. La presente tesi si focalizza inoltre sullo studio della modellizzazione dei materiali costituenti il moncone e sul contributo della compressione del liner sul moncone.

Analisi agli elementi finiti (FEM) di una protesi transfemorale sub-ischiatica ibrida : influenza delle caratteristiche meccaniche dei tessuti molli e contributo del liner

Bellomo, Silvia;Bigatti, Matteo
2019/2020

Abstract

This work aims to study the behaviour of the new type of socket designed, built and marketed by the company ITOP S.p.A. Officine Ortopediche: the hybrid sub-ischial socket. Using the Abaqus/CAE 6.14 software, the finite element analysis is carried out to calculate the stresses and deformations on the stump's external surface following the procedure of the donning of the prosthesis and the bipodalic stance in orthostatic posture. To verify the prosthesis's actual comfort, the stresses obtained in terms of contact pressure and shear stresses must fall within a range of values considered not harmful to the patient. This thesis focuses on studying the modelling of the materials constituting the stump and on the contribution of the liner compression on the stump.
MICCOLI, STEFANO
ING - Scuola di Ingegneria Industriale e dell'Informazione
28-apr-2021
2019/2020
Il presente lavoro si pone come obiettivo lo studio del comportamento della nuova tipologia di invasatura progettata, realizzata e commercializzata dall’azienda ITOP S.p.A. Officine Ortopediche: l’invasatura sub-ischiatica ibrida. Tramite l’utilizzo del software Abaqus/CAE 6.14 si conduce l’analisi agli elementi finiti, finalizzata al calcolo degli sforzi e delle deformazioni sulla superficie esterna del moncone sia al termine della calzata della protesi sia in posizione ortostatica con appoggio bipodalico. Al fine di verificare l’effettivo comfort della protesi, gli sforzi ottenuti in termini di pressione di contatto e di sforzi di taglio devono ricadere all’interno di un range di valori considerato non dannoso per il paziente. La presente tesi si focalizza inoltre sullo studio della modellizzazione dei materiali costituenti il moncone e sul contributo della compressione del liner sul moncone.
File allegati
File Dimensione Formato  
2021_04_Bellomo_Bigatti.pdf

accessibile in internet solo dagli utenti autorizzati

Descrizione: Testo della tesi
Dimensione 5.75 MB
Formato Adobe PDF
5.75 MB Adobe PDF   Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/173425