Open heart surgeries require deairing of the heart chambers to prevent embolization. CO2 catheters are being used to provide carbon dioxide at the point of insertion. To study the flow from catheters, it is necessary to design an appropriate setup. This study describes the procedure from obtaining CT scans, transforming them into 3D models and 3D printing them, to designing and building a rotocaster and creating a silicone mold from these models using this rotational caster. Different grades of silicone will be tested along with varying other factors that play a role in the final result. At the end, the advantages and disadvantages of each factor will be listed, and recommendations will be made to determine the best setup for creating artificial silicone hearts that are suitable for experimentation.

Gli interventi chirurgici a cuore aperto richiedono la degasazione delle camere cardiache per prevenire l'embolizzazione. I cateteri di CO2 vengono utilizzati per fornire anidride carbonica nel punto di inserimento. Per studiare il flusso dai cateteri, è necessario progettare una configurazione appropriata. Questo studio descrive la procedura dall'ottenimento di scansioni CT, trasformandole in modelli 3D e stampandole in 3D, fino alla progettazione e costruzione di un rotocaster e alla creazione di uno stampo in silicone da questi modelli utilizzando questa ruota rotante. Saranno testati diversi gradi di silicone insieme a vari altri fattori che giocano un ruolo nel risultato finale. Alla fine, verranno elencati i vantaggi e gli svantaggi di ciascun fattore e verranno formulate raccomandazioni per determinare la migliore configurazione per la creazione di cuori in silicone artificiale adatti alla sperimentazione.

Fabrication of anatomical models for experimentation : from CT scans to 3D silicone models

EL JURDI, JULIE
2020/2021

Abstract

Open heart surgeries require deairing of the heart chambers to prevent embolization. CO2 catheters are being used to provide carbon dioxide at the point of insertion. To study the flow from catheters, it is necessary to design an appropriate setup. This study describes the procedure from obtaining CT scans, transforming them into 3D models and 3D printing them, to designing and building a rotocaster and creating a silicone mold from these models using this rotational caster. Different grades of silicone will be tested along with varying other factors that play a role in the final result. At the end, the advantages and disadvantages of each factor will be listed, and recommendations will be made to determine the best setup for creating artificial silicone hearts that are suitable for experimentation.
DE GAETANO, FRANCESCO
ING - Scuola di Ingegneria Industriale e dell'Informazione
9-giu-2021
2020/2021
Gli interventi chirurgici a cuore aperto richiedono la degasazione delle camere cardiache per prevenire l'embolizzazione. I cateteri di CO2 vengono utilizzati per fornire anidride carbonica nel punto di inserimento. Per studiare il flusso dai cateteri, è necessario progettare una configurazione appropriata. Questo studio descrive la procedura dall'ottenimento di scansioni CT, trasformandole in modelli 3D e stampandole in 3D, fino alla progettazione e costruzione di un rotocaster e alla creazione di uno stampo in silicone da questi modelli utilizzando questa ruota rotante. Saranno testati diversi gradi di silicone insieme a vari altri fattori che giocano un ruolo nel risultato finale. Alla fine, verranno elencati i vantaggi e gli svantaggi di ciascun fattore e verranno formulate raccomandazioni per determinare la migliore configurazione per la creazione di cuori in silicone artificiale adatti alla sperimentazione.
File allegati
File Dimensione Formato  
Thesis-Final Draft.pdf

accessibile in internet per tutti

Descrizione: Thesis Report-Final Draft
Dimensione 2.76 MB
Formato Adobe PDF
2.76 MB Adobe PDF Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/176084