Ventricular assist devices (VADs) provide mechanical circulatory support for patients with advanced heart failure, but the non-physiological flow generated within rotary pumps may expose red blood cells to damaging mechanical stresses. This thesis evaluates the hydrodynamic and hemolytic behavior of two VADs, the centrifugal CentriMag and the axial-flow HeartMate II, through a combined in vitro and in silico approach. The in vitro study reconstructed pressure head--flow rate characteristics over the investigated operating ranges and performed bovine-blood hemolysis tests at selected working points. In parallel, computational fluid dynamics (CFD) simulations were developed to reproduce the experimental conditions, validate pressure-head generation, and interpret the internal pressure, velocity, and stress fields. Computational hemolysis was estimated using an Eulerian formulation and Lagrangian pathline-based models, including stress-based and strain-based formulations with two integration schemes. The CentriMag CFD model showed good hydraulic agreement with the in vitro data, with a total average pressure-head error of 5,0%. The HeartMate II model reproduced the expected axial-flow pump behavior, with a total average in vitro--in silico pressure-head error of 6,9%, although localized deviations remained at specific operating points. For both devices, the Eulerian MIH increased with rotational speed and decreased with flow rate, consistently with higher stress levels and shorter residence times. The Lagrangian analysis provided additional particle-history information, but also revealed sensitivity to pathline number, sampling tolerance, and the selected stress or strain formulation. Overall, the study demonstrates that pressure-head validation, flow-field analysis, and complementary Eulerian and Lagrangian hemolysis models can support the assessment of VAD hydraulic performance and hemolytic risk. Although further validation and model calibration are required, the proposed workflow provides a structured basis for integrating experimental testing and CFD in the evaluation of blood pumps.

I dispositivi di assistenza ventricolare (VAD) forniscono supporto circolatorio meccanico ai pazienti con insufficienza cardiaca avanzata; tuttavia, il flusso non fisiologico generato dalle pompe rotative può esporre i globuli rossi a sollecitazioni meccaniche dannose. Questa tesi valuta il comportamento idrodinamico ed emolitico di due VAD, la CentriMag centrifuga e l'HeartMate II assiale, mediante un approccio integrato in vitro e in silico. Lo studio in vitro ha ricostruito le curve prevalenza--portata negli intervalli operativi considerati e ha incluso prove di emolisi su sangue bovino in punti selezionati. In parallelo, simulazioni di fluidodinamica computazionale (CFD) sono state sviluppate per riprodurre le condizioni sperimentali, validare la generazione di prevalenza e interpretare i campi interni di pressione, velocità e stress. L'emolisi computazionale è stata stimata con una formulazione Euleriana e con modelli Lagrangiani basati su pathline, includendo formulazioni stress-based e strain-based con schemi HI3 e HI4. Il modello CFD della CentriMag ha mostrato un buon accordo idraulico con i dati in vitro, con un errore medio totale sulla prevalenza pari al 5,0%. Il modello HeartMate II ha riprodotto il comportamento atteso di una pompa assiale, con un errore medio totale in vitro--in silico del 6,9%, pur mantenendo scostamenti locali in alcuni punti operativi. Per entrambi i dispositivi, l'MIH Euleriano aumentava con la velocità di rotazione e diminuiva con la portata, coerentemente con stress più elevati e tempi di permanenza più brevi. L'analisi Lagrangiana ha aggiunto informazioni sulla storia delle particelle, ma ha evidenziato sensibilità al numero di pathline, alla tolleranza di campionamento e alla formulazione scelta. Nel complesso, lo studio mostra che validazione pressure-head, analisi dei campi di flusso e modelli emolitici Euleriani e Lagrangiani complementari possono supportare la valutazione delle prestazioni idrauliche e del rischio emolitico nei VAD. Pur richiedendo ulteriore validazione e calibrazione dei modelli, il workflow proposto offre una base strutturata per integrare prove sperimentali e CFD nello studio delle pompe ematiche.

Hydrodynamic and hemolytic evaluation of Ventricular Assist Devices: an in vitro and in silico study

Accetta, Matheus
2025/2026

Abstract

Ventricular assist devices (VADs) provide mechanical circulatory support for patients with advanced heart failure, but the non-physiological flow generated within rotary pumps may expose red blood cells to damaging mechanical stresses. This thesis evaluates the hydrodynamic and hemolytic behavior of two VADs, the centrifugal CentriMag and the axial-flow HeartMate II, through a combined in vitro and in silico approach. The in vitro study reconstructed pressure head--flow rate characteristics over the investigated operating ranges and performed bovine-blood hemolysis tests at selected working points. In parallel, computational fluid dynamics (CFD) simulations were developed to reproduce the experimental conditions, validate pressure-head generation, and interpret the internal pressure, velocity, and stress fields. Computational hemolysis was estimated using an Eulerian formulation and Lagrangian pathline-based models, including stress-based and strain-based formulations with two integration schemes. The CentriMag CFD model showed good hydraulic agreement with the in vitro data, with a total average pressure-head error of 5,0%. The HeartMate II model reproduced the expected axial-flow pump behavior, with a total average in vitro--in silico pressure-head error of 6,9%, although localized deviations remained at specific operating points. For both devices, the Eulerian MIH increased with rotational speed and decreased with flow rate, consistently with higher stress levels and shorter residence times. The Lagrangian analysis provided additional particle-history information, but also revealed sensitivity to pathline number, sampling tolerance, and the selected stress or strain formulation. Overall, the study demonstrates that pressure-head validation, flow-field analysis, and complementary Eulerian and Lagrangian hemolysis models can support the assessment of VAD hydraulic performance and hemolytic risk. Although further validation and model calibration are required, the proposed workflow provides a structured basis for integrating experimental testing and CFD in the evaluation of blood pumps.
ING - Scuola di Ingegneria Industriale e dell'Informazione
22-lug-2026
2025/2026
I dispositivi di assistenza ventricolare (VAD) forniscono supporto circolatorio meccanico ai pazienti con insufficienza cardiaca avanzata; tuttavia, il flusso non fisiologico generato dalle pompe rotative può esporre i globuli rossi a sollecitazioni meccaniche dannose. Questa tesi valuta il comportamento idrodinamico ed emolitico di due VAD, la CentriMag centrifuga e l'HeartMate II assiale, mediante un approccio integrato in vitro e in silico. Lo studio in vitro ha ricostruito le curve prevalenza--portata negli intervalli operativi considerati e ha incluso prove di emolisi su sangue bovino in punti selezionati. In parallelo, simulazioni di fluidodinamica computazionale (CFD) sono state sviluppate per riprodurre le condizioni sperimentali, validare la generazione di prevalenza e interpretare i campi interni di pressione, velocità e stress. L'emolisi computazionale è stata stimata con una formulazione Euleriana e con modelli Lagrangiani basati su pathline, includendo formulazioni stress-based e strain-based con schemi HI3 e HI4. Il modello CFD della CentriMag ha mostrato un buon accordo idraulico con i dati in vitro, con un errore medio totale sulla prevalenza pari al 5,0%. Il modello HeartMate II ha riprodotto il comportamento atteso di una pompa assiale, con un errore medio totale in vitro--in silico del 6,9%, pur mantenendo scostamenti locali in alcuni punti operativi. Per entrambi i dispositivi, l'MIH Euleriano aumentava con la velocità di rotazione e diminuiva con la portata, coerentemente con stress più elevati e tempi di permanenza più brevi. L'analisi Lagrangiana ha aggiunto informazioni sulla storia delle particelle, ma ha evidenziato sensibilità al numero di pathline, alla tolleranza di campionamento e alla formulazione scelta. Nel complesso, lo studio mostra che validazione pressure-head, analisi dei campi di flusso e modelli emolitici Euleriani e Lagrangiani complementari possono supportare la valutazione delle prestazioni idrauliche e del rischio emolitico nei VAD. Pur richiedendo ulteriore validazione e calibrazione dei modelli, il workflow proposto offre una base strutturata per integrare prove sperimentali e CFD nello studio delle pompe ematiche.
File allegati
File Dimensione Formato  
2026_07_Accetta_Tesi.pdf

solo utenti autorizzati a partire dal 02/07/2029

Descrizione: Tese
Dimensione 23.89 MB
Formato Adobe PDF
23.89 MB Adobe PDF   Visualizza/Apri
2026_07_Accetta_Executive_Summary.pdf

solo utenti autorizzati a partire dal 02/07/2029

Descrizione: Executive Summary
Dimensione 2.78 MB
Formato Adobe PDF
2.78 MB Adobe PDF   Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/260677