This study evaluates, through a Hospital-Based Health Technology Assessment (HB-HTA) approach, the potential introduction at the IRCCS Istituto Nazionale dei Tumori of Milan of an MR-guided adaptive radiotherapy pathway based on the MRIdian platform for selected patients with pancreatic adenocarcinoma, using the currently available CT/CBCT-guided pathway as the reference. The assessment was developed across the nine HTA dimensions by integrating scientific evidence, local operational data, economic modelling, and structured professional evaluations. The analysis considered clinical and technical-dosimetric aspects, operational workflows, process costs and budget impact, as well as organisational, equity, social, ethical, and legal implications. Results were ultimately integrated through Multi-Criteria Decision Analysis. Available evidence shows a relevant technical and dosimetric rationale for MRIdian, related to daily MR imaging, online plan adaptation, and intrafraction motion management, particularly in cases characterised by greater anatomical complexity and close proximity between the target and radiosensitive gastrointestinal organs. However, the available clinical evidence does not allow a direct comparative advantage of the MR-guided platform over the CT/CBCT-guided pathway to be established. Economic and organisational analyses also indicate greater resource use and higher complexity for the adaptive pathway, while equity and social, ethical, and legal aspects are strongly dependent on local implementation and access conditions. The multidimensional analysis shows that the value of MR-guided adaptive radiotherapy depends on the balance between potential benefits, sustainability, and appropriateness within the local context. Overall, the results do not support a generalised replacement of the TrueBeam pathway, but suggest a possible selective and complementary role for MRIdian/SMART in patients with greater anatomical and dosimetric complexity, within explicit selection criteria and a potentially multi-indication institutional strategy.

Il presente studio valuta, secondo un approccio di Hospital-Based Health Technology Assessment (HB-HTA), il possibile inserimento presso l’IRCCS Istituto Nazionale dei Tumori di Milano di un percorso di radioterapia adattiva MR-guided basato sulla piattaforma MRIdian, per pazienti selezionati con adenocarcinoma pancreatico, assumendo come riferimento il percorso CT/CBCT-guided attualmente disponibile. La valutazione è sviluppata secondo le nove dimensioni dell’HTA, integrando evidenze scientifiche, dati operativi locali, modellizzazione economica e valutazioni professionali strutturate. L’analisi considera gli aspetti clinici e tecnico-dosimetrici, i flussi operativi, i costi di processo e l’impatto sul budget, nonché le implicazioni organizzative, di equità, sociali, etiche e legali. I risultati sono stati infine integrati mediante Multi-Criteria Decision Analysis. Le evidenze disponibili mostrano per MRIdian un rilevante razionale tecnico-dosimetrico legato all’imaging RM quotidiano, all’adattamento online del piano e alla gestione intrafrazionale del movimento, particolarmente rilevanti nei casi a maggiore complessità anatomica e stretta prossimità tra bersaglio e organi gastrointestinali radiosensibili. Le evidenze cliniche non consentono tuttavia di attribuire un vantaggio comparativo diretto alla piattaforma MR-guided rispetto al percorso CT/CBCT-guided. L’analisi economica e organizzativa evidenzia inoltre un maggiore assorbimento di risorse e una maggiore complessità del percorso adattivo, mentre equità e aspetti sociali, etici e legali dipendono fortemente dalle condizioni locali di implementazione e accesso. L’analisi multidimensionale mostra che il valore della MR-guided adaptive radiotherapy dipende dal bilanciamento tra benefici potenziali, sostenibilità e appropriatezza nel contesto di riferimento. Nel complesso, i risultati non supportano una sostituzione generalizzata del percorso TrueBeam, ma suggeriscono per MRIdian/SMART un possibile ruolo selettivo e complementare nei pazienti a maggiore complessità anatomica e dosimetrica, nell’ambito di criteri espliciti di selezione e di una strategia aziendale potenzialmente multi-indicazione.

Dalla precisione al valore: valutazione Hospital-Based HTA della radioterapia adattiva MR-guided nel carcinoma pancreatico

PASCUCCI, GIULIANO
2025/2026

Abstract

This study evaluates, through a Hospital-Based Health Technology Assessment (HB-HTA) approach, the potential introduction at the IRCCS Istituto Nazionale dei Tumori of Milan of an MR-guided adaptive radiotherapy pathway based on the MRIdian platform for selected patients with pancreatic adenocarcinoma, using the currently available CT/CBCT-guided pathway as the reference. The assessment was developed across the nine HTA dimensions by integrating scientific evidence, local operational data, economic modelling, and structured professional evaluations. The analysis considered clinical and technical-dosimetric aspects, operational workflows, process costs and budget impact, as well as organisational, equity, social, ethical, and legal implications. Results were ultimately integrated through Multi-Criteria Decision Analysis. Available evidence shows a relevant technical and dosimetric rationale for MRIdian, related to daily MR imaging, online plan adaptation, and intrafraction motion management, particularly in cases characterised by greater anatomical complexity and close proximity between the target and radiosensitive gastrointestinal organs. However, the available clinical evidence does not allow a direct comparative advantage of the MR-guided platform over the CT/CBCT-guided pathway to be established. Economic and organisational analyses also indicate greater resource use and higher complexity for the adaptive pathway, while equity and social, ethical, and legal aspects are strongly dependent on local implementation and access conditions. The multidimensional analysis shows that the value of MR-guided adaptive radiotherapy depends on the balance between potential benefits, sustainability, and appropriateness within the local context. Overall, the results do not support a generalised replacement of the TrueBeam pathway, but suggest a possible selective and complementary role for MRIdian/SMART in patients with greater anatomical and dosimetric complexity, within explicit selection criteria and a potentially multi-indication institutional strategy.
ING - Scuola di Ingegneria Industriale e dell'Informazione
21-ott-2026
2025/2026
Il presente studio valuta, secondo un approccio di Hospital-Based Health Technology Assessment (HB-HTA), il possibile inserimento presso l’IRCCS Istituto Nazionale dei Tumori di Milano di un percorso di radioterapia adattiva MR-guided basato sulla piattaforma MRIdian, per pazienti selezionati con adenocarcinoma pancreatico, assumendo come riferimento il percorso CT/CBCT-guided attualmente disponibile. La valutazione è sviluppata secondo le nove dimensioni dell’HTA, integrando evidenze scientifiche, dati operativi locali, modellizzazione economica e valutazioni professionali strutturate. L’analisi considera gli aspetti clinici e tecnico-dosimetrici, i flussi operativi, i costi di processo e l’impatto sul budget, nonché le implicazioni organizzative, di equità, sociali, etiche e legali. I risultati sono stati infine integrati mediante Multi-Criteria Decision Analysis. Le evidenze disponibili mostrano per MRIdian un rilevante razionale tecnico-dosimetrico legato all’imaging RM quotidiano, all’adattamento online del piano e alla gestione intrafrazionale del movimento, particolarmente rilevanti nei casi a maggiore complessità anatomica e stretta prossimità tra bersaglio e organi gastrointestinali radiosensibili. Le evidenze cliniche non consentono tuttavia di attribuire un vantaggio comparativo diretto alla piattaforma MR-guided rispetto al percorso CT/CBCT-guided. L’analisi economica e organizzativa evidenzia inoltre un maggiore assorbimento di risorse e una maggiore complessità del percorso adattivo, mentre equità e aspetti sociali, etici e legali dipendono fortemente dalle condizioni locali di implementazione e accesso. L’analisi multidimensionale mostra che il valore della MR-guided adaptive radiotherapy dipende dal bilanciamento tra benefici potenziali, sostenibilità e appropriatezza nel contesto di riferimento. Nel complesso, i risultati non supportano una sostituzione generalizzata del percorso TrueBeam, ma suggeriscono per MRIdian/SMART un possibile ruolo selettivo e complementare nei pazienti a maggiore complessità anatomica e dosimetrica, nell’ambito di criteri espliciti di selezione e di una strategia aziendale potenzialmente multi-indicazione.
File allegati
File Dimensione Formato  
2026_10_Pascucci_Executive_Summary.pdf

accessibile in internet solo dagli utenti autorizzati

Dimensione 2.3 MB
Formato Adobe PDF
2.3 MB Adobe PDF   Visualizza/Apri
2026_10_Pascucci_Tesi.pdf

accessibile in internet solo dagli utenti autorizzati

Dimensione 13.54 MB
Formato Adobe PDF
13.54 MB Adobe PDF   Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/266418