Accurate measurement of intraocular pressure (IOP) is essential for the diagnosis and management of glaucoma. Although Goldmann applanation tonometry remains the clinical reference standard, its measurements are influenced by individual corneal biomechanical properties. This study aimed to evaluate an experimental tonometer prototype designed to estimate IOP while accounting for the biomechanical response of the cornea. The prototype was tested ex vivo on porcine eyes in which IOP was controlled using a hydrostatic pressure system. During corneal applanation, force and displacement measurements were used to calculate an elastic constant (K) describing the corneal mechanical response. The relationship between K and IOP was analysed using linear and quadratic fitting models, and the predictive performance of the prototype was evaluated on independent specimens. The estimated elastic constant showed a consistent inverse relationship with increasing IOP. Among the tested models, the linear approach provided the most robust and accurate pressure estimation, while the quadratic model did not offer a systematic improvement. Furthermore, the calibration strategy developed from one group of specimens successfully estimated IOP in an independent group, demonstrating the reproducibility of the proposed measurement principle. Overall, the results demonstrate the feasibility of a biomechanically compensated approach for estimating intraocular pressure and support further development of the prototype. Additional validation using larger datasets and human eyes will be required before clinical application.
La misurazione accurata della pressione intraoculare (IOP) è fondamentale per la diagnosi e il monitoraggio del glaucoma. Sebbene la tonometria ad applanazione di Goldmann rappresenti tuttora il metodo clinico di riferimento, le sue misurazioni sono influenzate dalle proprietà biomeccaniche individuali della cornea. Lo scopo di questo studio è stato valutare un prototipo sperimentale di tonometro progettato per stimare la IOP tenendo conto della risposta biomeccanica della cornea. Il prototipo è stato testato ex vivo su occhi porcini, nei quali la pressione intraoculare è stata controllata mediante un sistema di pressione idrostatica. Durante l’applanazione corneale sono stati misurati forza e spostamento, utilizzati per calcolare una costante elastica (K) rappresentativa della risposta meccanica della cornea. La relazione tra K e la IOP è stata analizzata mediante modelli di fitting lineare e quadratico, valutando successivamente la capacità del prototipo di stimare la pressione in campioni indipendenti. La costante elastica stimata ha mostrato una relazione inversa riproducibile con l’aumentare della IOP. Tra i modelli valutati, il fitting lineare ha fornito le stime più accurate e robuste, mentre il modello quadratico non ha determinato un miglioramento sistematico delle prestazioni. Inoltre, la strategia di calibrazione sviluppata su un primo gruppo di campioni ha consentito di stimare con buona accuratezza la IOP in un gruppo indipendente, dimostrando la riproducibilità del principio di misura proposto. Nel complesso, i risultati dimostrano la fattibilità di un approccio biomeccanicamente compensato per la stima della pressione intraoculare e supportano l’ulteriore sviluppo del prototipo. Saranno tuttavia necessari studi su un numero maggiore di campioni e la successiva validazione su occhi umani prima di una possibile applicazione clinica.
Calibration and experimental assessment of a prototype tonometer for intraocular pressure measurement
BACCALIN, COSTANZA;CAVARRETTA, SILVIA
2025/2026
Abstract
Accurate measurement of intraocular pressure (IOP) is essential for the diagnosis and management of glaucoma. Although Goldmann applanation tonometry remains the clinical reference standard, its measurements are influenced by individual corneal biomechanical properties. This study aimed to evaluate an experimental tonometer prototype designed to estimate IOP while accounting for the biomechanical response of the cornea. The prototype was tested ex vivo on porcine eyes in which IOP was controlled using a hydrostatic pressure system. During corneal applanation, force and displacement measurements were used to calculate an elastic constant (K) describing the corneal mechanical response. The relationship between K and IOP was analysed using linear and quadratic fitting models, and the predictive performance of the prototype was evaluated on independent specimens. The estimated elastic constant showed a consistent inverse relationship with increasing IOP. Among the tested models, the linear approach provided the most robust and accurate pressure estimation, while the quadratic model did not offer a systematic improvement. Furthermore, the calibration strategy developed from one group of specimens successfully estimated IOP in an independent group, demonstrating the reproducibility of the proposed measurement principle. Overall, the results demonstrate the feasibility of a biomechanically compensated approach for estimating intraocular pressure and support further development of the prototype. Additional validation using larger datasets and human eyes will be required before clinical application.| File | Dimensione | Formato | |
|---|---|---|---|
|
2026_10_Baccalin_Cavarretta_Tesi.pdf
non accessibile
Descrizione: tesi
Dimensione
24.86 MB
Formato
Adobe PDF
|
24.86 MB | Adobe PDF | Visualizza/Apri |
|
2026_10_Baccalin_Cavarretta_Executive Summary.pdf
non accessibile
Descrizione: executive summary
Dimensione
5.34 MB
Formato
Adobe PDF
|
5.34 MB | Adobe PDF | Visualizza/Apri |
I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
https://hdl.handle.net/10589/266595