The Dose Delivery System (DDS) at the CNAO Foundation utilizes ionization chambers (BOX1 and BOX2) that are sensitive to therapeutic intensities, on the order of 10^9 particles per second, however become less sensitive to low-intensity beams that are at times required in the CNAO Experimental Room (CNAO XPR). This thesis describes the final physical integration and experimental validation of a new low-intensity beam monitor specifically designed for single-particle detection. The detection system consists of two orthogonal planes of 128 scintillating fibers each, coupled with Hamamatsu Silicon Photomultiplier (SiPM) matrices. The data acquisition chain is built upon a CAEN DT5550W motherboard featuring a Xilinx Kintex-7 FPGA and an A55CIT4 piggyback board housing Citiroc 1A ASICs responsible for reading 128 SiPM channels. A custom firmware was developed using SCI-Compiler to implement a processing logic compatible with the DDS architecture, utilizing 16-bit counters, global latching mechanisms, and geometric channel remapping via ROM blocks to align the physical fiber positions with the output data stream. To bridge the communication gap between the CAEN electronics and the existing DDS hardware, the DDLowInt board interface board was employed for serial-to-parallel conversion, now integrated with a new optical fiber SerDes interface to ensure signal integrity and electrical isolation between the monitor and the DDS placed in the control room. The system was fully integrated into the CNAO Experimental Room control chain, where the DDS FPGA was programmed using LabVIEW to manage the data acquisition. The fully integrated system was validated in the CNAO Experimental Room (XPR) across five independent measurements using low-intensity carbon-ion beams. Finally, to support future carbon-ion arc therapy research --- where target volumes must rotate relative to a fixed beamline --- a custom target rotation stage based on a Newport SMC100 controller was integrated. The Ethernet-to-serial bridge was established and a control application was created using .NET assembly nodes. \\ This thesis successfully demonstrates the full commissioning of the Low intEnsity beam moNITOr (LENTO) and its integration into the CNAO Dose Delivery System. By providing real-time, sub-millimeter beam profile and fluence monitoring for single-particle regimes, LENTO bridges the low intensity monitoring gap previously present at CNAO.

Il Dose Delivery System (DDS) della Fondazione CNAO utilizza camere a ionizzazione (BOX1 e BOX2) che sono sensibili alle intensità terapeutiche, dell'ordine di 10^9 particelle al secondo, ma che tuttavia risultano meno sensibili ai fasci a bassa intensità talvolta richiesti nella Sala Sperimentale del CNAO (CNAO XPR). Questa tesi descrive l'integrazione fisica finale e la validazione sperimentale di un nuovo monitor per fasci a bassa intensità, progettato specificamente per il tracciamento della singola particella. Il sistema di rivelazione è composto da due piani ortogonali di 128 fibre scintillanti ciascuno, accoppiati a matrici di Fotomoltiplicatori al Silicio (SiPM) Hamamatsu. La catena di acquisizione dati si basa su una scheda madre CAEN DT5550W dotata di un FPGA Xilinx Kintex-7 e su una scheda mezzanine A55CIT4 che ospita gli ASIC Citiroc 1A, responsabili della lettura dei 128 canali SiPM. È stato sviluppato un firmware personalizzato tramite SCI-Compiler per implementare una logica di elaborazione compatibile con l'architettura del DDS, utilizzando contatori a 16 bit, meccanismi di latching globale e un riposizionamento geometrico dei canali via blocchi ROM, in modo da allineare le posizioni fisiche delle fibre con il flusso di dati in uscita. Per colmare il divario di comunicazione tra l'elettronica CAEN e l'hardware DDS esistente, è stata impiegata la scheda di interfaccia DDLowInt board per la conversione da seriale a parallelo, ora integrata con una nuova interfaccia SerDes a fibra ottica per garantire l'integrità del segnale e l'isolamento elettrico tra il monitor e il DDS. Il sistema è stato completamente integrato nella catena di controllo della Sala Sperimentale del CNAO (XPR), dove le FPGA del DDS sono state programmate in LabVIEW per gestire l'acquisizione dei dati. L'intero sistema integrato è stato validato nella Sala Sperimentale del CNAO attraverso cinque campagne di misura indipendenti, utilizzando fasci di ioni carbonio a bassa intensità. Infine, per supportare la ricerca futura sulla terapia ad arco con ioni carbonio --- in cui i volumi bersaglio devono ruotare rispetto a una linea di fascio fissa --- è stato integrato uno stadio di rotazione del bersaglio basato su un controllore Newport SMC100. È stato stabilito una connessione Ethernet-seriale ed è stata creata un'applicazione di controllo utilizzando nodi di assembly .NET. Questa tesi dimostra con successo il completo collaudo del monitor a fibre scintillanti Low intEnsity beam moNITOr (LENTO) e la sua integrazione nel Dose Delivery System del CNAO. Fornendo in tempo reale il profilo del fascio e il monitoraggio della fluenza con precisione sub-millimetrica nel regime di singola particella, LENTO colma il limite del monitoraggio a bassa intensità precedentemente presente al CNAO.

Experimental validation and system integration of a scintillating fiber monitor for low-intensity particle beam at CNAO

Morokhovska, Anastasiia
2025/2026

Abstract

The Dose Delivery System (DDS) at the CNAO Foundation utilizes ionization chambers (BOX1 and BOX2) that are sensitive to therapeutic intensities, on the order of 10^9 particles per second, however become less sensitive to low-intensity beams that are at times required in the CNAO Experimental Room (CNAO XPR). This thesis describes the final physical integration and experimental validation of a new low-intensity beam monitor specifically designed for single-particle detection. The detection system consists of two orthogonal planes of 128 scintillating fibers each, coupled with Hamamatsu Silicon Photomultiplier (SiPM) matrices. The data acquisition chain is built upon a CAEN DT5550W motherboard featuring a Xilinx Kintex-7 FPGA and an A55CIT4 piggyback board housing Citiroc 1A ASICs responsible for reading 128 SiPM channels. A custom firmware was developed using SCI-Compiler to implement a processing logic compatible with the DDS architecture, utilizing 16-bit counters, global latching mechanisms, and geometric channel remapping via ROM blocks to align the physical fiber positions with the output data stream. To bridge the communication gap between the CAEN electronics and the existing DDS hardware, the DDLowInt board interface board was employed for serial-to-parallel conversion, now integrated with a new optical fiber SerDes interface to ensure signal integrity and electrical isolation between the monitor and the DDS placed in the control room. The system was fully integrated into the CNAO Experimental Room control chain, where the DDS FPGA was programmed using LabVIEW to manage the data acquisition. The fully integrated system was validated in the CNAO Experimental Room (XPR) across five independent measurements using low-intensity carbon-ion beams. Finally, to support future carbon-ion arc therapy research --- where target volumes must rotate relative to a fixed beamline --- a custom target rotation stage based on a Newport SMC100 controller was integrated. The Ethernet-to-serial bridge was established and a control application was created using .NET assembly nodes. \\ This thesis successfully demonstrates the full commissioning of the Low intEnsity beam moNITOr (LENTO) and its integration into the CNAO Dose Delivery System. By providing real-time, sub-millimeter beam profile and fluence monitoring for single-particle regimes, LENTO bridges the low intensity monitoring gap previously present at CNAO.
ING - Scuola di Ingegneria Industriale e dell'Informazione
21-ott-2026
2025/2026
Il Dose Delivery System (DDS) della Fondazione CNAO utilizza camere a ionizzazione (BOX1 e BOX2) che sono sensibili alle intensità terapeutiche, dell'ordine di 10^9 particelle al secondo, ma che tuttavia risultano meno sensibili ai fasci a bassa intensità talvolta richiesti nella Sala Sperimentale del CNAO (CNAO XPR). Questa tesi descrive l'integrazione fisica finale e la validazione sperimentale di un nuovo monitor per fasci a bassa intensità, progettato specificamente per il tracciamento della singola particella. Il sistema di rivelazione è composto da due piani ortogonali di 128 fibre scintillanti ciascuno, accoppiati a matrici di Fotomoltiplicatori al Silicio (SiPM) Hamamatsu. La catena di acquisizione dati si basa su una scheda madre CAEN DT5550W dotata di un FPGA Xilinx Kintex-7 e su una scheda mezzanine A55CIT4 che ospita gli ASIC Citiroc 1A, responsabili della lettura dei 128 canali SiPM. È stato sviluppato un firmware personalizzato tramite SCI-Compiler per implementare una logica di elaborazione compatibile con l'architettura del DDS, utilizzando contatori a 16 bit, meccanismi di latching globale e un riposizionamento geometrico dei canali via blocchi ROM, in modo da allineare le posizioni fisiche delle fibre con il flusso di dati in uscita. Per colmare il divario di comunicazione tra l'elettronica CAEN e l'hardware DDS esistente, è stata impiegata la scheda di interfaccia DDLowInt board per la conversione da seriale a parallelo, ora integrata con una nuova interfaccia SerDes a fibra ottica per garantire l'integrità del segnale e l'isolamento elettrico tra il monitor e il DDS. Il sistema è stato completamente integrato nella catena di controllo della Sala Sperimentale del CNAO (XPR), dove le FPGA del DDS sono state programmate in LabVIEW per gestire l'acquisizione dei dati. L'intero sistema integrato è stato validato nella Sala Sperimentale del CNAO attraverso cinque campagne di misura indipendenti, utilizzando fasci di ioni carbonio a bassa intensità. Infine, per supportare la ricerca futura sulla terapia ad arco con ioni carbonio --- in cui i volumi bersaglio devono ruotare rispetto a una linea di fascio fissa --- è stato integrato uno stadio di rotazione del bersaglio basato su un controllore Newport SMC100. È stato stabilito una connessione Ethernet-seriale ed è stata creata un'applicazione di controllo utilizzando nodi di assembly .NET. Questa tesi dimostra con successo il completo collaudo del monitor a fibre scintillanti Low intEnsity beam moNITOr (LENTO) e la sua integrazione nel Dose Delivery System del CNAO. Fornendo in tempo reale il profilo del fascio e il monitoraggio della fluenza con precisione sub-millimetrica nel regime di singola particella, LENTO colma il limite del monitoraggio a bassa intensità precedentemente presente al CNAO.
File allegati
File Dimensione Formato  
2026_10_Morokhovska_ExecutiveSummary.pdf

accessibile in internet per tutti

Descrizione: Executive Summary Text
Dimensione 355.49 kB
Formato Adobe PDF
355.49 kB Adobe PDF Visualizza/Apri
2026_10_Morokhovska_Thesis.pdf

accessibile in internet per tutti

Descrizione: Thesis Text
Dimensione 19.38 MB
Formato Adobe PDF
19.38 MB Adobe PDF Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/266643