In the biomedical field, the use of techniques based on the analysis of light emitted by biological samples is becoming increasingly relevant for the study and characterization of different diseases. Among these techniques, Fluorescence Lifetime Imaging Microscopy (FLIM) provides information on the properties of biological tissues by analyzing the fluorescence decay time. FLIM relies on the principle of Time-Correlated Single-Photon Counting (TCSPC), through which the temporal distribution of detected photons is reconstructed by collecting a large number of events over time. One of the main limitations of TCSPC is the photon count rate, which must be kept sufficiently low with respect to the laser repetition rate to avoid pile-up effects and the consequent distortion of the measured temporal distribution. To overcome this limitation, a new measurement paradigm for TCSPC, referred to as ProActive, has been developed. It corrects pile-up distortion by compensating for undetected photons, taking into account, at each instant in time, the probability that the system is available to detect a new photon. This approach is based on an algorithm whose implementation requires high-throughput real-time processing and is therefore implemented on an FPGA. The algorithm was validated through a series of experiments conducted in the research laboratory, demonstrating its capability to extend TCSPC measurements to previously unexplored count rates, reaching several hundred Mevents/s. Such high count rates impose stringent requirements on the time-to-digital conversion stage, which must provide both high temporal resolution and ultra-fast event processing. In previous work, the algorithm was validated using a commercial event timer interfaced with an external FPGA, achieving a maximum throughput of 100 Mevents/s. While this setup successfully demonstrated the effectiveness of the proposed correction approach, the commercial event timer constitutes a bottleneck for applications requiring higher count rates. Furthermore, the need for dedicated commercial instrumentation and an external FPGA results in a relatively complex and costly system architecture. This thesis addresses these limitations by introducing a compact, fully FPGA-based TCSPC architecture. The proposed system integrates in the FPGA a shift register for event management and distribution, a multichannel TDC-based system for reconstructing the timing data required for TCSPC measurements, and the pile-up correction algorithm. The architecture achieves a dead time of only 1 ns, providing the high-throughput capability required to extend TCSPC measurements towards count rates in the order of hundreds of Mevents/s. The developed system was designed to support real-time data acquisition and visualization. A high-speed communication interface based on an FTDI module enables data transfer from the FPGA to a C# application, allowing the acquired measurements and temporal histograms to be monitored in real time. The proposed system combines the ultra-fast time-to-digital conversion capability of the TDC with real-time data processing based on a pile-up correction algorithm. This enables distortion-free TCSPC measurements while preserving the key advantages of the technique, including its high temporal resolution and photon-counting capability.

In ambito biomedico, l'impiego di tecniche basate sull'analisi della luce emessa da campioni biologici sta assumendo un'importanza sempre maggiore nello studio e nella caratterizzazione di diverse patologie. Tra queste, la Fluorescence Lifetime Imaging Microscopy (FLIM) consente di ottenere informazioni sulle proprietà dei tessuti biologici attraverso l'analisi del tempo di decadimento della fluorescenza. La tecnica FLIM si basa sul principio della Time-Correlated Single-Photon Counting (TCSPC), mediante il quale è possibile ricostruire la distribuzione temporale dei fotoni rilevati attraverso l'acquisizione di un elevato numero di eventi. Una delle principali limitazioni della TCSPC è rappresentata dal photon count rate, che deve essere mantenuto sufficientemente basso rispetto alla frequenza di ripetizione del laser al fine di evitare il fenomeno del pile-up e la conseguente distorsione della distribuzione temporale misurata. Per superare tale limitazione, è stato sviluppato un nuovo paradigma di misura per la TCSPC, denominato ProActive. Tale approccio consente di correggere la distorsione dovuta al pile-up compensando statisticamente i fotoni non rilevati, mediante la stima, in ogni istante temporale, della probabilità che il sistema sia disponibile per la rilevazione di un nuovo fotone. L'approccio si basa su un algoritmo la cui implementazione richiede un'elevata capacità di elaborazione dei dati in tempo reale e che, pertanto, viene implementato su FPGA. L'algoritmo è stato validato mediante una serie di esperimenti condotti in laboratorio, dimostrando la possibilità di estendere le misure TCSPC a count rate dell'ordine delle centinaia di Meventi/s. Il raggiungimento di tali valori richiede uno stadio di conversione tempo-digitale caratterizzato da un'elevata velocità di elaborazione degli eventi, oltre che da un'adeguata risoluzione temporale. In un precedente lavoro, la validazione dell'algoritmo è stata effettuata mediante un event timer commerciale interfacciato a una FPGA esterna, in grado di sostenere un throughput massimo di 100 Meventi/s. Sebbene tale configurazione abbia permesso di dimostrare l'efficacia dell'approccio di correzione proposto, l'event timer commerciale costituisce un limite per applicazioni caratterizzate da count rate superiori. Inoltre, l'impiego di strumentazione commerciale dedicata e di una FPGA esterna determina una soluzione complessa e poco compatta. La presente tesi affronta tali limitazioni attraverso lo sviluppo di un'architettura TCSPC compatta e interamente implementata su FPGA. Il sistema proposto integra in FPGA uno shift register per la gestione e distribuzione degli eventi, un sistema basato su TDC multicanale per la ricostruzione dei dati temporali per le misure di TCSPC e l'algoritmo di correzione del pile-up. L'architettura consente di eseguire all'interno dello stesso dispositivo sia la conversione tempo-digitale sia l'elaborazione dei dati in tempo reale. L'architettura sviluppata è caratterizzata da un dead time di solo 1 ns, rendendo possibile il supporto di elevati count rate e consentendo di estendere le misure TCSPC verso valori dell'ordine delle centinaia di Mevents/s. Il sistema è stato inoltre progettato per supportare l'acquisizione e la visualizzazione dei dati in tempo reale. Un'interfaccia di comunicazione ad alta velocità, basata su un modulo FTDI, consente il trasferimento dei dati dalla FPGA a un'applicazione sviluppata in C#, attraverso la quale è possibile monitorare in tempo reale le misure acquisite e i relativi istogrammi temporali. Il sistema proposto combina pertanto la conversione tempo-digitale ad elevata velocità del TDC con l'elaborazione real-time basata sull'algoritmo di correzione del pile-up. In questo modo, è possibile effettuare misure TCSPC prive della distorsione introdotta dal pile-up, preservando al contempo i principali vantaggi della tecnica, quali l'elevata risoluzione temporale e la capacità di rilevazione dei singoli fotoni.

Design and characterization of an FPGA-based multichannel TDC time-tagging architecture for real-time distortion-free TCSPC

Lotto, Giulio
2025/2026

Abstract

In the biomedical field, the use of techniques based on the analysis of light emitted by biological samples is becoming increasingly relevant for the study and characterization of different diseases. Among these techniques, Fluorescence Lifetime Imaging Microscopy (FLIM) provides information on the properties of biological tissues by analyzing the fluorescence decay time. FLIM relies on the principle of Time-Correlated Single-Photon Counting (TCSPC), through which the temporal distribution of detected photons is reconstructed by collecting a large number of events over time. One of the main limitations of TCSPC is the photon count rate, which must be kept sufficiently low with respect to the laser repetition rate to avoid pile-up effects and the consequent distortion of the measured temporal distribution. To overcome this limitation, a new measurement paradigm for TCSPC, referred to as ProActive, has been developed. It corrects pile-up distortion by compensating for undetected photons, taking into account, at each instant in time, the probability that the system is available to detect a new photon. This approach is based on an algorithm whose implementation requires high-throughput real-time processing and is therefore implemented on an FPGA. The algorithm was validated through a series of experiments conducted in the research laboratory, demonstrating its capability to extend TCSPC measurements to previously unexplored count rates, reaching several hundred Mevents/s. Such high count rates impose stringent requirements on the time-to-digital conversion stage, which must provide both high temporal resolution and ultra-fast event processing. In previous work, the algorithm was validated using a commercial event timer interfaced with an external FPGA, achieving a maximum throughput of 100 Mevents/s. While this setup successfully demonstrated the effectiveness of the proposed correction approach, the commercial event timer constitutes a bottleneck for applications requiring higher count rates. Furthermore, the need for dedicated commercial instrumentation and an external FPGA results in a relatively complex and costly system architecture. This thesis addresses these limitations by introducing a compact, fully FPGA-based TCSPC architecture. The proposed system integrates in the FPGA a shift register for event management and distribution, a multichannel TDC-based system for reconstructing the timing data required for TCSPC measurements, and the pile-up correction algorithm. The architecture achieves a dead time of only 1 ns, providing the high-throughput capability required to extend TCSPC measurements towards count rates in the order of hundreds of Mevents/s. The developed system was designed to support real-time data acquisition and visualization. A high-speed communication interface based on an FTDI module enables data transfer from the FPGA to a C# application, allowing the acquired measurements and temporal histograms to be monitored in real time. The proposed system combines the ultra-fast time-to-digital conversion capability of the TDC with real-time data processing based on a pile-up correction algorithm. This enables distortion-free TCSPC measurements while preserving the key advantages of the technique, including its high temporal resolution and photon-counting capability.
ING - Scuola di Ingegneria Industriale e dell'Informazione
21-ott-2026
2025/2026
In ambito biomedico, l'impiego di tecniche basate sull'analisi della luce emessa da campioni biologici sta assumendo un'importanza sempre maggiore nello studio e nella caratterizzazione di diverse patologie. Tra queste, la Fluorescence Lifetime Imaging Microscopy (FLIM) consente di ottenere informazioni sulle proprietà dei tessuti biologici attraverso l'analisi del tempo di decadimento della fluorescenza. La tecnica FLIM si basa sul principio della Time-Correlated Single-Photon Counting (TCSPC), mediante il quale è possibile ricostruire la distribuzione temporale dei fotoni rilevati attraverso l'acquisizione di un elevato numero di eventi. Una delle principali limitazioni della TCSPC è rappresentata dal photon count rate, che deve essere mantenuto sufficientemente basso rispetto alla frequenza di ripetizione del laser al fine di evitare il fenomeno del pile-up e la conseguente distorsione della distribuzione temporale misurata. Per superare tale limitazione, è stato sviluppato un nuovo paradigma di misura per la TCSPC, denominato ProActive. Tale approccio consente di correggere la distorsione dovuta al pile-up compensando statisticamente i fotoni non rilevati, mediante la stima, in ogni istante temporale, della probabilità che il sistema sia disponibile per la rilevazione di un nuovo fotone. L'approccio si basa su un algoritmo la cui implementazione richiede un'elevata capacità di elaborazione dei dati in tempo reale e che, pertanto, viene implementato su FPGA. L'algoritmo è stato validato mediante una serie di esperimenti condotti in laboratorio, dimostrando la possibilità di estendere le misure TCSPC a count rate dell'ordine delle centinaia di Meventi/s. Il raggiungimento di tali valori richiede uno stadio di conversione tempo-digitale caratterizzato da un'elevata velocità di elaborazione degli eventi, oltre che da un'adeguata risoluzione temporale. In un precedente lavoro, la validazione dell'algoritmo è stata effettuata mediante un event timer commerciale interfacciato a una FPGA esterna, in grado di sostenere un throughput massimo di 100 Meventi/s. Sebbene tale configurazione abbia permesso di dimostrare l'efficacia dell'approccio di correzione proposto, l'event timer commerciale costituisce un limite per applicazioni caratterizzate da count rate superiori. Inoltre, l'impiego di strumentazione commerciale dedicata e di una FPGA esterna determina una soluzione complessa e poco compatta. La presente tesi affronta tali limitazioni attraverso lo sviluppo di un'architettura TCSPC compatta e interamente implementata su FPGA. Il sistema proposto integra in FPGA uno shift register per la gestione e distribuzione degli eventi, un sistema basato su TDC multicanale per la ricostruzione dei dati temporali per le misure di TCSPC e l'algoritmo di correzione del pile-up. L'architettura consente di eseguire all'interno dello stesso dispositivo sia la conversione tempo-digitale sia l'elaborazione dei dati in tempo reale. L'architettura sviluppata è caratterizzata da un dead time di solo 1 ns, rendendo possibile il supporto di elevati count rate e consentendo di estendere le misure TCSPC verso valori dell'ordine delle centinaia di Mevents/s. Il sistema è stato inoltre progettato per supportare l'acquisizione e la visualizzazione dei dati in tempo reale. Un'interfaccia di comunicazione ad alta velocità, basata su un modulo FTDI, consente il trasferimento dei dati dalla FPGA a un'applicazione sviluppata in C#, attraverso la quale è possibile monitorare in tempo reale le misure acquisite e i relativi istogrammi temporali. Il sistema proposto combina pertanto la conversione tempo-digitale ad elevata velocità del TDC con l'elaborazione real-time basata sull'algoritmo di correzione del pile-up. In questo modo, è possibile effettuare misure TCSPC prive della distorsione introdotta dal pile-up, preservando al contempo i principali vantaggi della tecnica, quali l'elevata risoluzione temporale e la capacità di rilevazione dei singoli fotoni.
File allegati
File Dimensione Formato  
2026_10_Lotto_Tesi.pdf

accessibile in internet per tutti a partire dal 23/09/2027

Descrizione: Tesi
Dimensione 21.96 MB
Formato Adobe PDF
21.96 MB Adobe PDF   Visualizza/Apri
2026_10_Lotto_Executive_Summary.pdf

accessibile in internet per tutti a partire dal 23/09/2027

Descrizione: Executive Summary
Dimensione 962.57 kB
Formato Adobe PDF
962.57 kB Adobe PDF   Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/266675