Stroke frequently causes persistent upper-limb impairment, limiting independence in activities of daily living. Since many functional tasks require coordinated use of both arms, bimanual rehabilitation represents a promising strategy to promote motor recovery. This thesis develops an IMU-based master–slave framework for assisted bimanual upper-limb rehabilitation. The movement of the non-affected arm is acquired using three NGIMU sensors positioned on the trunk, upper arm and forearm, processed in real time, and used to reconstruct upper-limb kinematics and generate a mirrored contralateral reference for future exoskeleton control. The developed pipeline was preliminarily validated through isolated joint movements, functional reaching tasks and postural tests. Performance was assessed using range of motion, cross-talk, hand trajectory, movement duration, hand velocity, sagittal-plane distance, residual angular variation and acquisition rate. Results support the feasibility of the proposed approach and show repeatable three-dimensional motion reconstruction, while also identifying limitations related mainly to calibration and joint-angle decoupling. The system therefore provides a technological basis for future integration with a unilateral exoskeleton and subsequent validation in post-stroke rehabilitation.

L’ictus determina frequentemente deficit persistenti dell’arto superiore, con conseguente riduzione dell’autonomia nelle attività della vita quotidiana. Poiché molte azioni funzionali richiedono l’impiego coordinato di entrambi gli arti, la riabilitazione bimanuale rappresenta una strategia promettente per favorire il recupero motorio. La presente tesi sviluppa un framework master–slave basato su IMU per la riabilitazione bimanuale assistita dell’arto superiore. Il movimento dell’arto non affetto viene acquisito mediante tre sensori NGIMU posizionati su tronco, braccio e avambraccio, elaborato in tempo reale e utilizzato per ricostruire la cinematica dell’arto e generare un riferimento speculare controlaterale destinato al futuro controllo di un esoscheletro. La pipeline è stata validata preliminarmente mediante movimenti articolari isolati, task funzionali di reaching e test posturali. Le prestazioni sono state valutate attraverso range of motion, cross-talk, traiettoria della mano, durata del movimento, velocità, distanza dal piano sagittale, variazione angolare residua e frequenza di acquisizione. I risultati supportano la fattibilità dell’approccio e mostrano una ricostruzione tridimensionale sufficientemente ripetibile, evidenziando al contempo criticità legate soprattutto alla calibrazione e al disaccoppiamento articolare. Il sistema costituisce quindi una base tecnologica per la futura integrazione con un esoscheletro unilaterale e per successive validazioni in ambito post-ictus.

Assisted bimanual rehabilitation using IMU-based control: an early study

BONESI, GIULIA LILIANA
2025/2026

Abstract

Stroke frequently causes persistent upper-limb impairment, limiting independence in activities of daily living. Since many functional tasks require coordinated use of both arms, bimanual rehabilitation represents a promising strategy to promote motor recovery. This thesis develops an IMU-based master–slave framework for assisted bimanual upper-limb rehabilitation. The movement of the non-affected arm is acquired using three NGIMU sensors positioned on the trunk, upper arm and forearm, processed in real time, and used to reconstruct upper-limb kinematics and generate a mirrored contralateral reference for future exoskeleton control. The developed pipeline was preliminarily validated through isolated joint movements, functional reaching tasks and postural tests. Performance was assessed using range of motion, cross-talk, hand trajectory, movement duration, hand velocity, sagittal-plane distance, residual angular variation and acquisition rate. Results support the feasibility of the proposed approach and show repeatable three-dimensional motion reconstruction, while also identifying limitations related mainly to calibration and joint-angle decoupling. The system therefore provides a technological basis for future integration with a unilateral exoskeleton and subsequent validation in post-stroke rehabilitation.
ING - Scuola di Ingegneria Industriale e dell'Informazione
21-ott-2026
2025/2026
L’ictus determina frequentemente deficit persistenti dell’arto superiore, con conseguente riduzione dell’autonomia nelle attività della vita quotidiana. Poiché molte azioni funzionali richiedono l’impiego coordinato di entrambi gli arti, la riabilitazione bimanuale rappresenta una strategia promettente per favorire il recupero motorio. La presente tesi sviluppa un framework master–slave basato su IMU per la riabilitazione bimanuale assistita dell’arto superiore. Il movimento dell’arto non affetto viene acquisito mediante tre sensori NGIMU posizionati su tronco, braccio e avambraccio, elaborato in tempo reale e utilizzato per ricostruire la cinematica dell’arto e generare un riferimento speculare controlaterale destinato al futuro controllo di un esoscheletro. La pipeline è stata validata preliminarmente mediante movimenti articolari isolati, task funzionali di reaching e test posturali. Le prestazioni sono state valutate attraverso range of motion, cross-talk, traiettoria della mano, durata del movimento, velocità, distanza dal piano sagittale, variazione angolare residua e frequenza di acquisizione. I risultati supportano la fattibilità dell’approccio e mostrano una ricostruzione tridimensionale sufficientemente ripetibile, evidenziando al contempo criticità legate soprattutto alla calibrazione e al disaccoppiamento articolare. Il sistema costituisce quindi una base tecnologica per la futura integrazione con un esoscheletro unilaterale e per successive validazioni in ambito post-ictus.
File allegati
File Dimensione Formato  
Tesi_Bonesi.pdf

accessibile in internet per tutti

Dimensione 6.55 MB
Formato Adobe PDF
6.55 MB Adobe PDF Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/266920