The main purpose of the present work is the development and the realization of a polymeric coating in PCL on a substrate in Mg alloy AZ31 (3% aluminum, 1% zinc, 0.2% manganese) via dip coating technique for vascular stents. Coatings in PCL on Mg alloy AZ31 turn out to be continuous, homogeneous, smooth and with thickness in the range of 5-10 μm. Corrosion behavior of coated AZ31 alloy showed an improvement in corrosion resistance. However it was observed that coating’s imperfections persist along the edges of specimens, causing a rapid corrosion of the AZ31 substrate. It was also observed that the substrate's kinetics of corrosion strongly depends on grain size. In vitro indirect cytotoxicity tests were carried out for pure Mg, AZ31, AZ80 and ZM21 alloys. It was observed a decreasing in cell viability, probably due to high pH values.

Lo scopo principale del presente lavoro è lo sviluppo e la realizzazione di un rivestimento polimerico in PCL su un substrato in lega di Mg AZ31 (3% di alluminio, 1% di zinco, manganese 0,2%) tramite la tecnica di dip coating per applicazioni per stent vascolari. È stato progettato un dip coater, mediante il quale sono stati ottenuti rivestimenti in PCL sulla lega AZ31 continui, omogenei e con spessore nell’intorno di 5-10 μm. I risultati delle prove di degradazione hanno mostrato un miglioramento della resistenza alla corrosione della lega AZ31 rivestita con PCL rispetto alla lega non rivestita. Tuttavia è stato osservato che persistono imperfezioni del coating lungo i bordi dei provini, causando una rapida corrosione del substrato in AZ31. Inoltre è stato osservato che la cinetica di corrosione della lega AZ31 dipende fortemente dalla dimensione dei grani. Sono stati condotti test in vitro di citotossicità indiretta su Mg puro e sulle leghe AZ31,AZ80 e ZM21, durante i quali è stato osservato un calo della vitalità delle cellule, probabilmente a causa di elevati valori di pH.

Rivestimento in PCL per stent biodegradabili in lega di Mg : ottimizzazione del processo e studio della degradazione

ARLATI, ELISABETTA
2011/2012

Abstract

The main purpose of the present work is the development and the realization of a polymeric coating in PCL on a substrate in Mg alloy AZ31 (3% aluminum, 1% zinc, 0.2% manganese) via dip coating technique for vascular stents. Coatings in PCL on Mg alloy AZ31 turn out to be continuous, homogeneous, smooth and with thickness in the range of 5-10 μm. Corrosion behavior of coated AZ31 alloy showed an improvement in corrosion resistance. However it was observed that coating’s imperfections persist along the edges of specimens, causing a rapid corrosion of the AZ31 substrate. It was also observed that the substrate's kinetics of corrosion strongly depends on grain size. In vitro indirect cytotoxicity tests were carried out for pure Mg, AZ31, AZ80 and ZM21 alloys. It was observed a decreasing in cell viability, probably due to high pH values.
FARE', SILVIA
MIGLIAVACCA, FRANCESCO
ING II - Scuola di Ingegneria dei Sistemi
20-dic-2012
2011/2012
Lo scopo principale del presente lavoro è lo sviluppo e la realizzazione di un rivestimento polimerico in PCL su un substrato in lega di Mg AZ31 (3% di alluminio, 1% di zinco, manganese 0,2%) tramite la tecnica di dip coating per applicazioni per stent vascolari. È stato progettato un dip coater, mediante il quale sono stati ottenuti rivestimenti in PCL sulla lega AZ31 continui, omogenei e con spessore nell’intorno di 5-10 μm. I risultati delle prove di degradazione hanno mostrato un miglioramento della resistenza alla corrosione della lega AZ31 rivestita con PCL rispetto alla lega non rivestita. Tuttavia è stato osservato che persistono imperfezioni del coating lungo i bordi dei provini, causando una rapida corrosione del substrato in AZ31. Inoltre è stato osservato che la cinetica di corrosione della lega AZ31 dipende fortemente dalla dimensione dei grani. Sono stati condotti test in vitro di citotossicità indiretta su Mg puro e sulle leghe AZ31,AZ80 e ZM21, durante i quali è stato osservato un calo della vitalità delle cellule, probabilmente a causa di elevati valori di pH.
Tesi di laurea Magistrale
File allegati
File Dimensione Formato  
2012_Dicembre_Arlati.pdf

non accessibile

Descrizione: Testo della tesi
Dimensione 7.79 MB
Formato Adobe PDF
7.79 MB Adobe PDF   Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/72548