Neointimal Hyperplasia (NIH) and In-Stent Restenosis (ISR), i.e. the re-narrowing of a blood vessel after stent implantation, remains the main limiting factor of Percutaneous Coronary Interventions (PCI), known as most popular clinical approach for the treatment of coronary artery diseases. Aim of this work is to use Computational Fluid Dynamics tools to analyze and compare multiple factors that may affect ISR. In particular we will investigate oxygen and drug transfer phenomena in a realistic geometry of porcine coronary artery, obtained by micro-CT techniques. The results of computational simulations will be compared with histological images obtained at 14 days after stent implantation in the animal model.

L'iperplasia della neointima (NIH) e la conseguente ristenosi (ISR), ovvero il restringimento del lume vascolare in seguito all'impianto di uno stent, costituiscono i principali fattori limitanti degli interventi coronarici percutanei (Percutaneous Coronary Interventions, PCI), noti in ambito clinico quali principali trattamenti chirurgici per le patologie occlusive allo stadio avanzato a carico delle arterie coronarie. Mediante l'utilizzo di strumenti di Fluidodinamica Computazionale, il presente lavoro di tesi si prefigge come obiettivo l’analisi di diversi fattori che possono influenzare la ISR. In particolare, verrà investigato il trasporto di ossigeno e di farmaco in una geometria realistica di arteria coronaria porcina ottenuta attraverso micro-CT. I risultati delle simulazioni al calcolatore saranno poi confrontati con immagini istologiche ottenute a 14 giorni dall’impianto dello stent nel modello animale.

Trasporto di massa in arterie coronarie trattate con Stent. Analisi computazionale nella piattaforma ANSYS

CAPUTO, MARCO;CIANCIOLO, CLAUDIA
2011/2012

Abstract

Neointimal Hyperplasia (NIH) and In-Stent Restenosis (ISR), i.e. the re-narrowing of a blood vessel after stent implantation, remains the main limiting factor of Percutaneous Coronary Interventions (PCI), known as most popular clinical approach for the treatment of coronary artery diseases. Aim of this work is to use Computational Fluid Dynamics tools to analyze and compare multiple factors that may affect ISR. In particular we will investigate oxygen and drug transfer phenomena in a realistic geometry of porcine coronary artery, obtained by micro-CT techniques. The results of computational simulations will be compared with histological images obtained at 14 days after stent implantation in the animal model.
ZUNINO, PAOLO
CHIASTRA, CLAUDIO
CUTRI', ELENA
ING II - Scuola di Ingegneria dei Sistemi
20-dic-2012
2011/2012
L'iperplasia della neointima (NIH) e la conseguente ristenosi (ISR), ovvero il restringimento del lume vascolare in seguito all'impianto di uno stent, costituiscono i principali fattori limitanti degli interventi coronarici percutanei (Percutaneous Coronary Interventions, PCI), noti in ambito clinico quali principali trattamenti chirurgici per le patologie occlusive allo stadio avanzato a carico delle arterie coronarie. Mediante l'utilizzo di strumenti di Fluidodinamica Computazionale, il presente lavoro di tesi si prefigge come obiettivo l’analisi di diversi fattori che possono influenzare la ISR. In particolare, verrà investigato il trasporto di ossigeno e di farmaco in una geometria realistica di arteria coronaria porcina ottenuta attraverso micro-CT. I risultati delle simulazioni al calcolatore saranno poi confrontati con immagini istologiche ottenute a 14 giorni dall’impianto dello stent nel modello animale.
Tesi di laurea Magistrale
File allegati
File Dimensione Formato  
2012_12_Caputo_Cianciolo.pdf

accessibile in internet per tutti

Descrizione: Testo della tesi
Dimensione 19.4 MB
Formato Adobe PDF
19.4 MB Adobe PDF Visualizza/Apri

I documenti in POLITesi sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10589/72707